826

From Sensoray Technical Wiki
Revision as of 10:59, 25 April 2013 by Admin (Talk | contribs)

Jump to: navigation, search

Calibration errors

On 826 SDKs earlier than version 3.2.0, analog calibration values will not be applied without J6 (labeled "Calibration Enable") installed. A missing shunt is intended to protect against accidental overwriting of calibration values, but in these SDKs it also prevents the reading of those values. This is resolved in SDK version 3.2.0 and above; in these versions the shunt functions as intended and must be installed only when calibrating the board (though leaving it installed all the time is okay).

If board calibration is incorrect, make sure J6 is installed or upgrade to SDK version 3.2.0 or higher. The 826 SDK can be downloaded from the 826 product page.

Personal tools
Namespaces

Variants
Actions
Toolbox